• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AISYS CS2; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 02/01/2018
Event Type  malfunction  
Manufacturer Narrative
The distributor performed a checkout of the equipment and confirmed the reported complaint.The breathing circuit was cleaned, and flow and pressure calibration was performed.No report of patient involvement.(b)(4).The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that, during the preoperative checkout, mechanical ventilation was not functioning as expected.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7302529
MDR Text Key101171992
Report Number2112667-2018-00434
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K132530
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial
Report Date 02/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE H10
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/01/2018
Initial Date FDA Received02/28/2018
Date Device Manufactured02/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-