• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 7.5MM X 45MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 7.5MM X 45MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3505-7545
Device Problem Disassembly (1168)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2018-00223.
 
Event Description
It was reported that two pedicle screws disassembled during final tightening within surgery.They were removed and replaced with an alternative screw to complete the procedure.There were no reports of patient impacts associated with this event.This is report two of two for this event.
 
Manufacturer Narrative
The returned screw was examined.The screw components were found to have disassembled due to deformation on the swivel rings, which was caused when the screw was locked into position and then unlocked and repositioned during the procedure.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED POLYAXIAL SCREW 7.5MM X 45MM
Type of Device
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7302530
MDR Text Key101277143
Report Number3012447612-2018-00224
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK132884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3505-7545
Device Lot Number87NJ
Other Device ID Number(01)00889024335776(10)87NJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-