• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-33J
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 02/13/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device or a picture of the alleged defect was not provided at the time of this report.The device history record (dhr) has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.The dhr shows that the device was assembled and inspected according to our specifications.Customer complaint cannot be confirmed based only on the information provided.It is necessary to have the physical sample in order to perform a proper and thorough investigation to confirm the alleged defect and determine a root cause.If the sample becomes available, this investigation will be updated with the evaluation results.
 
Event Description
Customer complaint alleges the adaptor was unstable and couldnot be connected to the flow meter properly.Alleged issue reported as prior to patient use.No report of patient involvement.
 
Event Description
Customer complaint alleges the adaptor was unstable and could not be connected to the flow meter properly.Alleged issue reported as prior to patient use.No report of patient involvement.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and it was observed that there was damage on the thread of the adaptor.Functional testing was also performed and the sample passed the tests.During the setup of the oxygen entrainment testing, however, it was observed that the assembly of the nut adaptor and the upper body component was unstable.Even with that condition, the sample was able to be tested on oxygen entrainment testing but the testing failed.Based on the functional inspection results of the sample received, the complaint is confirmed.Although the condition observed on the sample provided, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The damage found on the thread of the adaptor is most likely caused by the end user during the connection of the adaptor into the flowmeter that causes an unstable connection issue.Personnel of the assembly line were notified on (b)(6) 2018 for awareness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
MDR Report Key7302969
MDR Text Key101572065
Report Number3004365956-2018-00060
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date06/19/2022
Device Catalogue Number031-33J
Device Lot Number74F1700608
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2017
Date Manufacturer Received04/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-