ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number EL5ML
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.
 
Event Description
It was reported that during a laparoscopic unknown procedure, the clip fed into the jaws with the legs bending inward.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Batch # p93d7j.Device analysis: the analysis results found that the el5ml device was received with no damage in the external components.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired and the clips did not advance into the jaw.The device was noted to have the tip of the advancer bent.The instrument was disassembled, upon disassembly the advancer was confirmed to be bent and 3 clips were found inside clip track.A possible cause for the condition of bent advancer is actuating the device when the jaws are not clear of the trocar.Actuating the device with the jaws closed may bend the advancer.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key7306702
MDR Text Key101600139
Report Number3005075853-2018-08274
Device Sequence Number1
Product Code FZP
UDI-Device Identifier20705036001840
UDI-Public20705036001840
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Catalogue NumberEL5ML
Device Lot NumberP4RY8N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2018
Date Manufacturer Received03/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-