Catalog Number 03260763160 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/10/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was requested for return, but was not provided.
|
|
Event Description
|
The customer complained of false positive nitrite results for 3 patient samples that were tested on a urisys 1100 urine analyzer with serial number (b)(4).Patient 1 initial nitrite result from the meter was positive.The doctor looked at the urine under a microscope and "did not see any nitrites".On (b)(6) 2018 patient 2 initial nitrite result from the meter was positive.The doctor sent out a urine culture out of the laboratory and the nitrite result was negative.On (b)(6) 2018 patient 3 initial nitrite result from the meter was positive.The doctor looked at the urine under a microscope and "did not see any nitrites".The microscopic observation and culture testing was performed using the same urine sample that was used for the strip measurement.The erroneous results were reported outside of the laboratory to a doctor.There was no treatment or medication provided to the patients.There was no allegation of an adverse event.The patients are in stable condition.The urine samples were not swirled and the urine was dropped onto the strip using a pipette.The testing was performed immediately after collecting the urine samples.The test strips were stored properly.Control testing was not recently performed on the meter.The analyzer and test strips were requested for investigation.Only the meter was returned for investigation.Retention material of lot #20318400 was measured on an iu urisys 1100 with native urine and a nitrite-dilution-series.The results fulfilled the requirements.No false positive results were observed.
|
|
Manufacturer Narrative
|
The urine samples were dropped onto the test strip using a pipette and according to product labeling the test strips are suppose to be dipped into the urine sample.Further investigation also noted that nitrites can not be observed using a microscope.Nitrite producing bacteria may be potentially seen under a microscope but not all bac types produce nitrites.
|
|
Manufacturer Narrative
|
Since qc testing was not performed with the meter, an instrument issue cannot be excluded.Further investigation could not be performed as the suspect product was not returned.
|
|
Manufacturer Narrative
|
The investigation was unable to find a definitive root cause.
|
|
Search Alerts/Recalls
|