Model Number 173016
Device Problem Detachment Of Device Component (1104)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/29/2017
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during a procedure.The needle came loose and broke off into the patient.The needle is still in the patient.There was patient involvement and medical intervention was required.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Device was returned loaded with a broken needle in left jaw of device.During testing and examination by engineering it was determined that there was no evidence of any malfunction of the device.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly or component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO STITCH
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middletown ave.
north haven, CT 06473
2034925563
MDR Report Key7309913
MDR Text Key101389552
Report Number9612501-2018-00291
Device Sequence Number1
Product Code OCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model Number173016
Device Catalogue Number173016
Device Lot NumberJ7K1923EX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/01/2018
Initial Date FDA Received03/02/2018
Supplement Dates Manufacturer Received03/19/2018
04/12/2018
05/09/2018
Supplement Dates FDA Received04/04/2018
04/16/2018
05/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-