On (b)(6) 2015 a field service engineer (fse) was dispatched.When fse arrived, low pressure error came up again and valve 3 had no pressure.The fse replaced the valve 3 and primed but there was air in the line and fse found buffer 3 straw in bag curled up and too long.The fse primed all and ran patient samples to prime.Pressure was good with no errors.Preventive maintenance was done the following day.Low pressure problem was still there.The fse replaced the check valves; pressure came right up.Precision, calibrators, qc, etc.All were good and all passed within acceptable range.The g8 instrument was functioning as intended and errors cleared.The customer was able to resume operations.No further action was required by fse.The probable cause of the event was related to the check valves.A complaint history review for serial number (b)(4) was conducted from (b)(6) 2014 through (b)(6) 2015 for similar complaints.There were no similar complaints identified during the search period.(b)(4) is submitting on behalf of the foreign manufacturer tosoh corporation, per exemption number (b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).
|