Model Number WCD 4000
Device Problem Invalid Sensing (2293)
Patient Problem No Information (3190)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (does not detect electrodes) was confirmed.As received, the belt would not communicate with a test monitor.Upon evaluation, the 3.3v wire was open inside the trunk cable.The cause of the inability to detect the electrodes is the open wire.The root cause of the open wire is excessive force placed on the cable.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor contacted zoll to report that a patient's device was not detecting the electrodes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key7314622
MDR Text Key101617940
Report Number3008642652-2018-02114
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 03/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/08/2018
Initial Date FDA Received03/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-