Model Number G8
Device Problems Complete Blockage (1094); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/16/2016
Event Type  malfunction  
Manufacturer Narrative
Additional narrative: this report is being submitted due to a retrospective review conducted under capa (b)(4).(b)(4).Device evaluation by manufacturer: the field service engineer (fse) was dispatched on (b)(6) 2016.The fse replaced the large syringe tip and found crystal build up on the rotor seal, replaced the seal.The engineer cleaned the dried buffer from inside the instrument, rebooted the instrument and allowed the system to do a complete initialization.The fse ran 38 samples with no issues.The customer reportedly ran calibrators and controls and patients with no issues.The g8 was working as intended.No further action required by fse.The probable cause of the event was a clogged rotor seal.
 
Event Description
On (b)(6) 2016 a customer reported high total areas (tas) with their g8 analyzer.The customer was unable to run samples and requested service to address the reported event.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay reporting hba1c patient results.There was no indication of any patient intervention or adverse event consequence due to the delay of reporting patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8, PRODUCT CODE: LCP
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7319829
MDR Text Key101997957
Report Number8031673-2018-01452
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/07/2018
Distributor Facility Aware Date03/16/2016
Device Age16 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/07/2018
Initial Date Manufacturer Received 03/16/2016
Initial Date FDA Received03/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-