• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Pressure Problem (3012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2015
Event Type  malfunction  
Manufacturer Narrative
On (b)(6) 2015 a field service engineer (fse) was dispatched.The fse replaced both check valves on pump and adjusted flow factor (ff), which resolved the issue and all errors cleared.The g8 analyzer was functioning as intended.No further action required by field service.The probable cause of the event was related to the check valves.A complaint history review for (b)(4) was conducted from (b)(6) 2014 through (b)(6) 2015 for similar complaints.There was one (1) similar complaints identified during the search period.(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).
 
Event Description
On (b)(6) 2015 a customer reported a peak pattern issue with the g8 analyzer.A field service engineer (fse) was dispatched.The customer is unable to run hbaic patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting of hbaic patient results.There was no indication of any patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
TOSOH HLC-723G8 ANALYZER G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7321358
MDR Text Key102076062
Report Number8031673-2018-01829
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/07/2018
Distributor Facility Aware Date10/26/2015
Device Age8 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/07/2018
Date Manufacturer Received10/26/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-