• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number E100-29M
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
An event of a piece of the valve holder breaking and traveling to the patient's leg was reported.The results of the investigation are inconclusive since the device was not returned for analysis, however four photos were received for analysis.Based solely on the aforementioned photos from the field, the plug portion of the valve holder appeared to have become dislodged from the valve holder.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
In (b)(6) 2018, a 29mm epic valve was implanted.During implant, a piece of the valve holder detached and was not recognized by the implanting medical team or the implanting md.The patient recovered and was discharged without any problems.At an unknown point following discharge, while in another country, the patient developed symptoms and evaluation identified a foreign material in the patient's leg.Intervention was taken to retrieve the foreign material from the patient's leg and it is reported the piece was consistent the white plug portion of the snap fit valve holder.The valve remains implanted and the detached holder piece has no impact on valve functionality.Patient identifier, age, and weight are protected under local privacy laws, and therefore are not recorded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC STENTED PORCINE HEART VALVE W/FLEXFIT SYSTEM
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key7325201
MDR Text Key101950199
Report Number3001883144-2018-00028
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P040021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/03/2021
Device Model NumberE100-29M
Device Catalogue NumberE100-29M
Device Lot Number6145188
Other Device ID Number05414734027533
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/12/2018
Initial Date FDA Received03/08/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-