Catalog Number 367342
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2018
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: unknown.Device manufacture date: unknown.Investigation: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Additionally, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.As no samples or photos were received for evaluation, the customer's indicated failure mode was not observed by bd.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the safety of a bd vacutainer® winged safety push button blood collection set ¿broke off¿ during use.There was no report of exposure, injury or medical intervention.
 
Manufacturer Narrative
Bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for a broken and missing iv needle retraction button with the incident lot was observed.Additionally, the samples were observed to be heavily damaged, the iv needles were retracted and the samples were missing the male luer.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Based on evaluation of the customer samples, the customer¿s indicated failure mode for a broken and missing iv needle retraction button with the incident lot was observed.Based on the investigation, a root cause could not be determined.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® WINGED SAFETY PUSH BUTTON BLOOD COLLECTION SET
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
MDR Report Key7327524
MDR Text Key102199989
Report Number1024879-2018-00066
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903673426
UDI-Public50382903673426
Combination Product (y/n)N
PMA/PMN Number
K030573
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 04/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367342
Device Lot NumberUNKNOWN
Initial Date Manufacturer Received 02/14/2018
Initial Date FDA Received03/09/2018
Supplement Dates Manufacturer Received02/14/2018
Supplement Dates FDA Received04/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-