The hill-rom technician found all four brake casters needed to be replaced.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the brake casters to determine if the bed moves when you activate the brake, replace or adjust when necessary.Check the tires for cuts, wear, tread life, etc.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in january 2018.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced all four brake casters to resolve the issue.Based on this information, no further action is required.
|