• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE TOTALCARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE TOTALCARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1900M006970
Device Problems Device Operates Differently Than Expected (2913); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/26/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found all four brake casters needed to be replaced.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the brake casters to determine if the bed moves when you activate the brake, replace or adjust when necessary.Check the tires for cuts, wear, tread life, etc.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in january 2018.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced all four brake casters to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the bed would not stay in place when the brakes were locked and pressure was applied to the bed.The bed was located in the bed shop at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
lori daulton
1069 state route 46 east
batesville, IN 47006
8129312280
MDR Report Key7331112
MDR Text Key102205727
Report Number1824206-2018-00112
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Service Personnel
Type of Report Initial
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1900M006970
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/26/2018
Initial Date FDA Received03/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-