Catalog Number 07.02077.001
Device Problem Bent (1059)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.
 
Event Description
It was reported that a lenke probe was found bent during a routine set inspection.There were no surgical or patient impacts associated with this issue.
 
Manufacturer Narrative
The returned lumbar lenke probe was evaluated.The evaluation showed that the tip is bent.The cause of failure cannot be determined at this time as the conditions of use when the device failed are unknown.Bending may occur due to dense bone quality of the patient, off-axis use of the probe, attempting to alter the trajectory of the probe within the vertebrae.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDICLE PROBE - STRAIGHT LUMBAR LENKE
Type of Device
PROBE
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7331371
MDR Text Key102211271
Report Number3012447612-2018-00252
Device Sequence Number1
Product Code HXB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number07.02077.001
Device Lot Number63041036
Other Device ID Number(01)00889024009998(10)63041036
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/28/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received09/07/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/02/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-