(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).On (b)(6) 2016, a field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse found that a low pressure error occurred on the first quality control (qc) and sample.The fse also found there was a new reagent buffer that was added overnight.To ensure sure that there was no air in the lines, the fse primed all of the buffers and the wash then ran the qc.All errors cleared and the g8 instrument is operational.No further action is required by field service.The most probable cause for error 101 pressure low was air in lines due to empty buffers.
|
On (b)(6) 2016, a customer reported error 101 pressure low on the g8 instrument.The customer stated that they replaced buffer bags on the instrument.Technical support advised the customer to ensure all buffer bags are full and to perform a "remove air from pump procedure" in the maintenance section of the operators manual.The customer is unable to run patient samples on hba1c diabetes assay.On (b)(6) 2016, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|