• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 11CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 11CM ANGLE ATTACHMENT; MOTOR, DRILL, ELECTRIC - ATTACHMENT Back to Search Results
Catalog Number QD11
Device Problems Overheating of Device (1437); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure it was observed that the angle attachment device was heating up and was inoperable.It was not reported if there was a delay in the procedure due to the event.It was not reported if a spare device was available for use.There was patient involvement reported.There were no patient or user injuries reported.It was reported there was no medical intervention or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the locking components and bearings were damaged.The device also failed pretest for thimble lock operation, nose tube assembly and cutter insertion.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Upon further investigation of the device evaluation, it was determined that the device controls were blocked, the bearing and sleeve were damaged and the glue joint to sleeve was loose.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11CM ANGLE ATTACHMENT
Type of Device
MOTOR, DRILL, ELECTRIC - ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key7333490
MDR Text Key102212139
Report Number1045834-2018-50490
Device Sequence Number1
Product Code HBC
UDI-Device Identifier845384005926
UDI-Public(01)845384005926(11)150518
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberQD11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/13/2018
Initial Date FDA Received03/13/2018
Supplement Dates Manufacturer Received04/26/2018
06/27/2018
Supplement Dates FDA Received05/03/2018
07/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-