Model Number P1900Q007678
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technical support representative suggested that the account check the nurse cable.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Check for current leakage at the nurse call system communication connections.Warning: a communication cable must be used for beds that have nurse call.Failure to do could cause patient injury.For beds that have nurse call systems, a communication cable must be connected between the bed and facility communication system.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.Three attempts have been made regarding a resolution to this contact line, with no response.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the nurse call was not working at the nurse's station, but was working on the bed.The bed was located at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTALCARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key7333510
MDR Text Key102213978
Report Number1824206-2018-00113
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Hospital Service Technician
Type of Report Initial
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP1900Q007678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/06/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-