DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS Back to Search Results
Catalog Number 221750041
Device Problems Failure to Disconnect (2541); Device-Device Incompatibility (2919); Naturally Worn (2988)
Patient Problem Not Applicable (3189)
Event Date 02/14/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Surgeon impacted the cup and could not unscrew the impactor handle from cup he had to remove the cup.Another surgeon came in and also tried to unscrew cup from handle, but could not do it.The surgeon had to open new instr set to get another impactor cup handle and had to ream to next size as first cup removal did affect cup size.All this did not cause any harm to patient but delayed case for about 30 minutes.I have tried impactor handle on another dni cup i have and it seems ok.It looks that the threads that were damaged during impaction were the ones in the cup but unfortunately, sterilization did not keep the cup after they finally unscrewed it from the handle with pliers.I can return impactor handle but no cup (b)(6).(b)(4).One ea 221750041 one ea 121722052 ps: the same thing happened on (b)(6), (b)(4), in a different hospital,same size cup but different lot #.Patient consequence? no.Is the information being submitted for this complaint all the details that are known/available regarding this event? yes.
 
Manufacturer Narrative
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN STRAIGHT CUP IMPACTOR
Type of Device
HIP INSTRUMENTS : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7337248
MDR Text Key102470471
Report Number1818910-2018-55021
Device Sequence Number1
Product Code MRA
UDI-Device Identifier10603295098980
UDI-Public10603295098980
Combination Product (y/n)N
PMA/PMN Number
P070026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 02/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number221750041
Device Lot NumberNB132437
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Date Manufacturer Received01/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-