DEPUY ORTHOPAEDICS, INC. 1818910 PINN STRAIGHT CUP IMPACTOR; HIP INSTRUMENTS : IMPACTORS
|
Back to Search Results |
|
Catalog Number 221750041 |
Device Problems
Failure to Disconnect (2541); Device-Device Incompatibility (2919); Naturally Worn (2988)
|
Patient Problem
Not Applicable (3189)
|
Event Date 02/14/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Surgeon impacted the cup and could not unscrew the impactor handle from cup he had to remove the cup.Another surgeon came in and also tried to unscrew cup from handle, but could not do it.The surgeon had to open new instr set to get another impactor cup handle and had to ream to next size as first cup removal did affect cup size.All this did not cause any harm to patient but delayed case for about 30 minutes.I have tried impactor handle on another dni cup i have and it seems ok.It looks that the threads that were damaged during impaction were the ones in the cup but unfortunately, sterilization did not keep the cup after they finally unscrewed it from the handle with pliers.I can return impactor handle but no cup (b)(6).(b)(4).One ea 221750041 one ea 121722052 ps: the same thing happened on (b)(6), (b)(4), in a different hospital,same size cup but different lot #.Patient consequence? no.Is the information being submitted for this complaint all the details that are known/available regarding this event? yes.
|
|
Manufacturer Narrative
|
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|