On (b)(6) 2016, the customer reported high, then low pressure, followed by "filter clogged popped tubing" with their g7 analyzer.Technical support (ts) advised the customer to replace the filter, reconnect the tubing, and repeat the patient sample.Ts advised the customer to call back if the issue persisted.On (b)(6) 2016, the customer contacted ts to report that she had replaced the filter but noted that the peak tubing was broken and would not stay attached.On (b)(6) 2016, field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|