• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER Back to Search Results
Model Number G8
Device Problem Partial Blockage (1065)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/28/2015
Event Type  malfunction  
Manufacturer Narrative
A field service engineer arrived at the customer site on 12/30/2015 to the address the reported event.The fse found an object inside the loader and removed a wood stick,cleaned and lubricated the sample loader and performed loader alignment verification.Several sample (x10) were run to verify the instrument, ran controls and ran 10 patient samples.Quality controls (qc) were acceptable.Chromatograms also passed.The instrument is working good now,the service is completed and no further action required by field service.The most probable cause of the reported event was due to a sample loader stuck due to a fallen object.(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).
 
Event Description
On (b)(6) 2015 a customer reported a syringe error 706 a bent sample needle with the g8.The customer was unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7339924
MDR Text Key103144780
Report Number8031673-2018-01684
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/14/2018
Distributor Facility Aware Date12/28/2015
Device Age7 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/14/2018
Initial Date Manufacturer Received 12/28/2015
Initial Date FDA Received03/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-