MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED HUMIDIFICATION CHAMBER; AUTOFEED CHAMBER Back to Search Results
Model Number MR290V
Device Problem Filling Problem (1233)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) method: the complaint mr290v vented autofeed humidification chamber was returned to fisher and paykel healthcare in (b)(4) for investigation.The chamber was visually inspected and functionally tested.Results: visual inspection revealed no damage to the chamber.The floats were found to be moving freely.During the functional test the water level of both primary and secondary floats was checked.Both floats operated correctly and the water level attained was within specification.Conclusion: we were unable to determine the cause of the reported issue, as the complaint mr290 chamber was found to be operating correctly.All mr290 chambers have float function and valve testing performed twice during production.A failure of any test would result in rejection of the chamber before distribution.The user instructions that accompany the mr290 state the following: set appropriate ventilator alarms.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.
 
Event Description
A hospital in (b)(6) reported via a distributor that "water overflowed" from an mr290v humidification chamber during patient use.No patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED HUMIDIFICATION CHAMBER
Type of Device
AUTOFEED CHAMBER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
sharon kuruvilla
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key7340502
MDR Text Key102595172
Report Number9611451-2018-00199
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial
Report Date 02/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290V
Device Catalogue NumberMR290V
Device Lot Number170911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-