• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977A260
Device Problems Break (1069); Device Operates Differently Than Expected (2913); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/12/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from an hcp via a manufacture representative regarding a patient implanted with an implantable neurostimulator (ins).It was reported that during implant, the hcp stated that the stylet felt like something was 'chewed up' inside the lead and they felt a 'gritty fee' when removing the stylet.The hcp was subsequently unable to fully re-insert the original stylet or the other stylets in the kit.The hcp decided not to use this lead as they were unable to steer the lead effectively and they used another lead instead, which worked fine.The issue was resolved at the time of the call.The lead was to be sent back to the manufacture for analysis.No patient complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Device was returned for analysis.No additional information was provided.
 
Manufacturer Narrative
Analysis of the 977a260 lead ((b)(4)) indicated that no anomalies were identifies.The lead passed all testing in the laboratory.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRIS SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7341033
MDR Text Key102460712
Report Number2649622-2018-02714
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00643169109391
UDI-Public00643169109391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2022
Device Model Number977A260
Device Catalogue Number977A260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/12/2018
Initial Date FDA Received03/15/2018
Supplement Dates Manufacturer Received03/13/2018
04/25/2018
Supplement Dates FDA Received04/04/2018
05/09/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
-
-