A medtronic representative went to the site to test the equipment.The representative was unable to replicate the reported issue.It was reported that evaluation of the polaris spectra system control unit (scu) error logs showed that communication could not be established through the main cable.The cable was subsequently replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The cable was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
|
A site representative reported that, while in a spinal fusion, the navigation system displayed a localizer not connected message in the application software.After multiple restarts of the navigation system, functionality was restored.There was a reported delay to the procedure of thirty minutes due to this issue.There was no impact on patient outcome.No additional information was provided.
|