• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Increase in Pressure (1491)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).On 30-jun-2017, fse visited the customer to address the reported event.During servicing, fse found increased pressure that was caused by an aging column.The column was replaced and the flow rate was adjusted.Fse confirmed that the pressure was in range.The analyzer was operating as expected.There was no further action required by fse.The most probable cause of the reported event was due to overuse of column filter.
 
Event Description
On (b)(6) 2017, a customer reported the pressure was on the high end of acceptability with their g8 analyzer.Additionally, the retention were elevated.Technical support advised the customer to adjust the flow rate, power, warmup and run primes.The retention times returned to normal but the pressure remained elevated.Technical support suggested that the customer that the customer replace the column.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7344242
MDR Text Key102600947
Report Number8031673-2018-02842
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/16/2018
Distributor Facility Aware Date06/29/2017
Device Age2 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/16/2018
Date Manufacturer Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Congenital Anomaly;
-
-