Device 1 of 2 reference mfr report#1627487-2018-02434 it was reported during replacement of the scs system (reference mfr.Report: 1627487-2018-02143), the distal tip end of the second lead separated when the physician was attempting to remove it.Reportedly, the broken lead was left inside the patient and the remaining system devices were explanted and replaced as scheduled.Postoperatively, effective therapy was established with the new system.Note: both leads are being reported as it is unknown which lead was broken.
|