The tosoh field service engineer (fse) went to the customer site to evaluate the g8 analyzer and perform preventive maintenance (pm) on the instrument.The fse found that no cycle count was entered and the instrument had reached over 23,000 cycles.Following pm, operation of the g8 analyzer was verified.Calibration, quality controls and precision testing were performed and all were within acceptable range.In addition, the fse set the flag for 20,000 cycles to alert the customer to schedule the next pm.The most potential causes of the high pressure may be related to malfunction of the grad sensor and/or a clogged filter or column.Review of the g8 variant analysis mode operator's manual provides the following information: the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.When the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: (a) greater than the pressure on the column inspection report + 4 mpa, then replace the filter.(b) less than the pressure on the column inspection report, then proceed with priming the column.Yearly maintenance is due at 20,000 cycles.Columns are warranted for 2500 injections.Change filter after 400 injections.(b)(4).This report is being submitted due to a retrospective review conducted under (b)(4).
|