• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE FAIRFIELD UNKNOWN; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Catalog Number UNKNOWN
Device Problems Disconnection (1171); Pumping Stopped (1503)
Patient Problems Death (1802); Ventricular Fibrillation (2130)
Event Date 02/10/2018
Event Type  malfunction  
Manufacturer Narrative
The device was not returned and could not be evaluated.It was discarded by the user.We are unable to confirm the reported event.If new information becomes available, a supplemental report will be submitted.(b)(4).
 
Event Description
It was reported that while delivering intra-aortic balloon (iab) therapy on a patient, the patient coded and went into ventricular fibrillation.During this event, the iab pump lost ecg trigger and stopped pumping.This occurred when the customer was attempting to transfer the patient.The patient expired during this event.There was no reported malfunction of the iab.The patient's death is not being attributed to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
DATASCOPE FAIRFIELD
15 law drive
fairfield NJ 07004
Manufacturer Contact
15 law drive
fairfield, NJ 07004
MDR Report Key7348159
MDR Text Key103015019
Report Number2248146-2018-00174
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeYR
Date Manufacturer Received02/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Death;
-
-