Model Number N/A |
Device Problems
Difficult to Insert (1316); Difficult To Position (1467)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated the product is in the process of being returned for investigation.Once the investigation has been completed, a follow up mdr will be submitted.Concomitant medical products: item (b)(4), g7 pps ltd acet shell 54f, lot 3878790.
|
|
Event Description
|
It was reported that the surgeon could not seat the liner into the g7 cup.Surgeon verified liner was proper match for the cup, but was unable to seat after several attempts.No adverse event was reported regarding the patient.Additional information was requested and received.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned device revealed gouging around the scallops of the liner and circular indentations were found on the outer radius that indicate a screw came in contact with the liner during attempts at impaction.No dimensional analysis will be performed due to the impaction of the liner.Device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|