Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated the product is in the process of being returned for investigation.Once the investigation has been completed, a follow up mdr will be submitted.Concomitant medical products: item (b)(4), g7 pps ltd acet shell 54f, lot 3878790.
 
Event Description
It was reported that the surgeon could not seat the liner into the g7 cup.Surgeon verified liner was proper match for the cup, but was unable to seat after several attempts.No adverse event was reported regarding the patient.Additional information was requested and received.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned device revealed gouging around the scallops of the liner and circular indentations were found on the outer radius that indicate a screw came in contact with the liner during attempts at impaction.No dimensional analysis will be performed due to the impaction of the liner.Device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL E1 LINER
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7355359
MDR Text Key103157932
Report Number0001825034-2018-02041
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/05/2019
Device Model NumberN/A
Device Catalogue Number010000935
Device Lot Number3467512
Other Device ID Number(01) 00880304527102
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/09/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-