MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 595000-001
Device Problems Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm while supporting the patient.The customer also reported that the patient switched to the backup freedom driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
Visual inspection of the driver's internal components revealed a broken white wire and frayed insulation on the red wire of the flow sensor to the main printed circuit board assembly (pcba) cable.The alarm history was reviewed and revealed one recorded alarm - a 22 fault code, correlating with a sudden change in voltage detected (fluctuations detected in a time frame of less than or equal to 1 second) during operation of the driver.This alarm was produced by the malfunction of the airflow sensor cable, which was found to be severed, and was the cause of the customer-reported alarm.Syncardia has an open corrective and preventative action (capa) that determined the assembly routing was causing the air flow sensor cable wires to become pinched between two internal driver components during or following driver assembly.This pinching appears to precipitate a severing of the wire.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7358321
MDR Text Key103292397
Report Number3003761017-2018-00096
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Remedial Action Replace
Type of Report Initial,Followup
Report Date 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/13/2018
Initial Date FDA Received03/21/2018
Supplement Dates Manufacturer Received03/13/2018
Supplement Dates FDA Received08/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age16 YR
-
-