MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Display Difficult to Read (1181); Image Display Error/Artifact (1304)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 03/06/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the issue was observed when the companion 2 driver was not supporting a patient.The companion 2 driver has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The companion 2 driver was not supporting a patient.The customer, a syncardia certified hospital, reported that upon receipt of the companion 2 driver, the display screen was dark and barely visible.
 
Manufacturer Narrative
The customer-reported "dark and barely visible" companion 2 driver display was confirmed and reproduced during incoming functional testing of the driver where the display was observed to be very dim and hard to read.The root cause of the dark screen was determined to be due to the partial connection of the companion 2 lcd inverter cable to the inverter board on the driver's lcd display, as observed during internal inspection of the driver.Investigation testing determined that all of the components functioned properly, and the driver passed all functional testing.Due to the properly functioning components, it is suspected that the cable was not inserted properly during the initial assembly of the driver.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7358324
MDR Text Key103292259
Report Number3003761017-2018-00092
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2018
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-