Model Number 397002-001
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2018
Event Type  malfunction  
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.The customer-reported issue of the driver's compressor cycling on and off was confirmed.Investigation testing determined the root cause to be a malfunction of the manual pressure regulator.Syncardia has a corrective and preventive action (capa) to address the issue of compressor cycling.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressor cycled on and off while the driver was connected to wall air and supporting a patient.The customer also reported that the patient was subsequently switched to a back-up driver.There was no reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7358325
MDR Text Key103143639
Report Number3003761017-2018-00085
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Remedial Action Replace
Type of Report Initial
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age60 YR
-
-