Device evaluated by mfr: visual inspection revealed dried body fluids on the handle, main body tubing and on the distal end.The distal end has twisted spiral marks in the main body tubing.Rings 1, 2 & 3 were partially flattened by some sort of tool.Tool marks can be seen on rings 2 & 3.Body fluid found under the edge of ring 1.Electrical continuity checks revealed no electrical shorts.Electrical opens were found for ring 1, ring 2 and magnetic sensor.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.X-ray found twisted electrical wires and a broken ring 1 wire.The distal end was dissected.Ring 1 and ring 2 wires were broken at the rings.As viewed from the end of the tip, the electrical wires, center support, steering wires and irrigation tube had been severely twisted in the counter-clockwise direction.The magnetic sensor insulation was dissected.The sensor wire was broken.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is user / use error as there was an act or omission of an act that resulted in a different medical device response than intended by the manufacturer or expected by the user, due to torsional overstress and tooling marks.(b)(4).
|