• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE; NEBULIZER (DIRECT PATIENT INTE Back to Search Results
Catalog Number 031-33J
Device Problem Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 03/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device or a picture of the alleged defect was not provided at the time of this report.The device history record (dhr) of the device has been reviewed and no issues or discrepancies were found which could potentially relate to this complaint.Dhr shows that the product was assembled and inspected according to our specifications.It is necessary to receive the physical sample in order to perform a proper and thorough investigation.No conclusion can be established at this time.Customer complaint cannot be confirmed.If the sample becomes available this report will be updated with the evaluation results.However, the personnel of the assembly line were notified for awareness.
 
Event Description
Customer complaint alleges "that the connecting part of flow meter and adaptor was unstable, then it was replaced by a new kit.The new one was used without any issue".
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.The sample was returned for evaluation.A visual exam was performed and it was observed that the nut adaptor did not spin freely.Functional testing was also performed and the sample passed the tests.During the setup of the oxygen entrainment testing, however, it was observed that the assembly of the nut adaptor and the upper body component was unstable.Even with that condition, the sample was able to be tested with no functional issues.After the testing was finished, the adaptor was carefully disassembled from the upper body and it was visually inspected.During the visual inspection it was found there was wear on the internal tabs.Attempts to duplicate the failure mode were performed and there are two ways to duplicate them: the first one is by overtightening the nut adaptor into the flow meter.The second one is by manipulating the assembly connection.(con't) other remarks: based on the investigation performed, the reported complaint was confirmed.Although the condition observed on the sample provided, there is not sufficient evidence to assure that this issue was originated during the manufacturing assembly or molding process.The wear on the internal tabs of the adaptor most likely was caused by the end user during the connection of the adaptor into the flowmeter.However, the personnel of the assembly line were notified on apr-11-2018 for awareness.
 
Event Description
Customer complaint alleges "that the connecting part of flow meter and adaptor was unstable, then it was replaced by a new kit.The new one was used without any issue".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON NEBULIZER ADAPTOR 033,STERILE,JAPANESE
Type of Device
NEBULIZER (DIRECT PATIENT INTE
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key7361331
MDR Text Key103298935
Report Number3004365956-2018-00084
Device Sequence Number1
Product Code CAF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K153010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date07/24/2022
Device Catalogue Number031-33J
Device Lot Number74G1702387
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-