• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G141
Device Problems Device Alarm System (1012); High impedance (1291); Low impedance (2285)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/13/2018
Event Type  malfunction  
Manufacturer Narrative
At this time, the product remains in service.If additional information is received, this report will be updated.
 
Event Description
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) was being checked, as the patient heard beeping.The beeping was in response to the left ventricular (lv) lead impedance measurement, that was less than 200 ohms.The pace impedance reading during the visit, was stable, and there were no sensing issues.Reportedly, the lv lead impedance had always trended daily low and was sometimes dipping just below 200. the field representative turned off the audible beeping for the lv lead. no further actions were taken.No adverse patient effects were reported.
 
Event Description
Additional information was reported which indicated that the beeping was in response to the right ventricular (rv) shock lead impedance measurement that was greater than 200 ohms.There was no issue with the lv lead.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7362054
MDR Text Key103281413
Report Number2124215-2018-05209
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534553
UDI-Public00802526534553
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/17/2017
Device Model NumberG141
Other Device ID NumberINOGEN CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 02/13/2018
Initial Date FDA Received03/22/2018
Supplement Dates Manufacturer Received03/23/2018
Supplement Dates FDA Received04/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
-
-