Boston scientific received information that this local representative contacted boston scientific's technical services.The local representative was informed by the clinic nurse who reported a right atrial (ra) impedance alert from this patient's latitude home monitoring system of greater than 3000 ohms.A review of stored data showed nominal ra impedances of approximately 550 ohms with intermittent spikes to greater than 3000 ohms.Additionally, a review of stored data identified atrial tachycardia episodes that shows minute ventilation (mv) rrt-noise on the ra electrogram.The mfv rrt-noise was caused by high impedance.The high impedance could be due to a setscrew connection or primary non-boston scientific lead issue.Since the implant of this non-boston scientific ra lead is relatively new (2016), the technical service's consultant suspect that the root cause may be due to a connection issue.The technical service's consultant recommended that be be seen in-clinic and reprogram the respiratory-related trends off.Diagnostic testing should be performed on the ra lead during patient isometrics and pocket manipulation.Subsequently, boston scientific received information that this local representative contacted boston scientific's technical services.The local representative discussed this event from an historical perspective.The technical service's consultant reported that the respiratory related trends is programmed off.Patient isometrics were performed with the patient's left arm but no electrogram noise occurred.The physician has elected to continue monitoring the ra pacing impedance.All indicators point to a spring contact issue with the non-boston scientific ra lead.
|