BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ADAPTER WITH LUER ADAPTER; BLOOD COLLECTION SET
|
Back to Search Results |
|
Catalog Number 367300 |
Device Problems
Connection Problem (2900); Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown/.
|
|
Event Description
|
It was reported that before use a bd vacutainer® adapter with luer adapter was found with foreign matter as ¿white glue was on the screw part and customer couldn't connect to holder.¿ there was no report of exposure, injury or medical intervention needed.
|
|
Manufacturer Narrative
|
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos and samples were evaluated and the customer¿s indicated failure mode for foreign matter with the incident lot was observed.Evaluation of the customer samples was performed and excess epoxy was observed on the product.However, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Investigation conclusion: based on evaluation of the customer photos and samples, the customer¿s indicated failure mode for foreign matter on the product with the incident lot was observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
|
|
Search Alerts/Recalls
|
|
|