• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ADAPTER WITH LUER ADAPTER; BLOOD COLLECTION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ADAPTER WITH LUER ADAPTER; BLOOD COLLECTION SET Back to Search Results
Catalog Number 367300
Device Problems Connection Problem (2900); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/05/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown/.
 
Event Description
It was reported that before use a bd vacutainer® adapter with luer adapter was found with foreign matter as ¿white glue was on the screw part and customer couldn't connect to holder.¿ there was no report of exposure, injury or medical intervention needed.
 
Manufacturer Narrative
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos and samples were evaluated and the customer¿s indicated failure mode for foreign matter with the incident lot was observed.Evaluation of the customer samples was performed and excess epoxy was observed on the product.However, bd was unable to determine the specific lot number associated with this complaint.Therefore, a review of the device history record could not be conducted.Investigation conclusion: based on evaluation of the customer photos and samples, the customer¿s indicated failure mode for foreign matter on the product with the incident lot was observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ADAPTER WITH LUER ADAPTER
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
MDR Report Key7363278
MDR Text Key103428938
Report Number1024879-2018-00098
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
PMA/PMN Number
K921520
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 04/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number367300
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/12/2018
Date Manufacturer Received03/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-