• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723GA ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723GA ANALYZER G8 Back to Search Results
Model Number G8
Device Problems Device Issue (2379); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/01/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).On (b)(6) 2016, a field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse spoke with the customer and advised him to check the end of the drain tube that goes into the drain into the floor.The customer identified a buildup of debris.The fse advise the customer to clear the debris.The g8 instrument is operational.No further action is required by field service.The most probable cause for the leak was a clogged drain tube.
 
Event Description
On (b)(6) 2016, a customer reported a leak on the g8 instrument.The customer stated that liquid was on the counter but not in the pan.The customer is unable to run patient samples on hba1c diabetes assay.On (b)(6) 2016, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723GA ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
ste. 101
south san francisco, CA 94080
6506368123
MDR Report Key7365385
MDR Text Key103523316
Report Number8031673-2018-03576
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/23/2018
Distributor Facility Aware Date03/01/2016
Device Age8 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/23/2018
Date Manufacturer Received03/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-