Model Number G8
Device Problems Device Issue (2379); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).On (b)(6) 2017, a field service engineer (fse) was dispatched to the customer's facility to address the reported event.The fse found that the left to x axis was off slightly.The fse adjusted the whole loader to the left.The fse also rerouted the needle tube.The fse ran patient samples with no z1 axis errors.The customer later stated that the error continued.The fse ran several tests and did not find an error.The fse found that the alignment was off for the y-sample.The fse made the adjustment and realigned the x1.The fse lubricated the z axis with triflow and ran patient samples with no z1 errors.All errors cleared and the g8 instrument is operational.No further action is required by field service.The most probable cause for error 710 z1 axis was the xaxis needed an adjustment.
 
Event Description
On (b)(6) 2017, a customer reported error 710 z1 axis with the g8 analyzer.The customer is unable to run hba1c patient samples.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723GA ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
ste. 101
south san francisco, CA 94080
6506368123
MDR Report Key7365392
MDR Text Key103725849
Report Number8031673-2018-03586
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/23/2018
Distributor Facility Aware Date05/09/2017
Device Age11 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/23/2018
Initial Date Manufacturer Received 05/09/2017
Initial Date FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-