Catalog Number V041163000000
Device Problem Device Stops Intermittently (1599)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.Multiple unsuccessful attempts were made to obtain the patient outcome.
 
Event Description
In this event it was reported that a vdw silver reciproc motor was suddenly stopping.The event outcome is unknown as of this mdr evaluation.
 
Manufacturer Narrative
Evaluation found a battery defect (voltage breakdown) and the motor contained oil/fluid inside.Both were replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY VDW GMBH
bayerwaldstrasse 15
munich, 81737
GM  81737
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
helen lewis
221 w. philadelphia st.
suite 60w
york, PA 17401
7178494229
MDR Report Key7366881
MDR Text Key103700191
Report Number9611053-2018-00033
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 04/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041163000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/29/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-