• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).Submission of this report does not constitute an admission that the importer or manufacturer's product caused or contributed to the event.On (b)(4) 2017, fse conducted follow-up with the customer over-the-phone to address the reported event.Fse followed up with the customer after the tubing been replaced to confirm functionality of the device.No further issues with leaking were reported.No further action was required by field service.The most probable cause of the reported event was due to a clogged prefilter.
 
Event Description
On (b)(6) 2017, the customer reported leaking with their g8 analyzer after preventative maintenance.Technical support (ts) instructed the customer to replace the prefilter.On (b)(4) 2017, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba- koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7368236
MDR Text Key103681506
Report Number8031673-2018-04235
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/23/2018
Distributor Facility Aware Date05/25/2017
Device Age6 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/23/2018
Initial Date Manufacturer Received 05/25/2017
Initial Date FDA Received03/23/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-