(b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).On (b)(4) 2016, an fse initially conducted a service call with the customer to troubleshoot the problem.During the service call, fse advised the customer to perform a drain flush but while doing so, the customer found a leak on the injection valve.The customer then reported that the high pressures were present constantly.Fse then visited the customer to address the reported event.During the evaluation, fse found the 1-6 line ferrules were crushed and were screwed on too tightly.When the fse replaced the 1-6 line, the pressure dropped to normal range.Patient samples and controls were ran with acceptable results.The instrument was verified as operational.There was no further action required by fse.The most probable cause of the reported event was peek tubing overtightened.
|
On (b)(6) 2016, a customer reported getting high pressures with their g8 analyzer.The customer reports that they changed the 1-6 tubing and the peek tubing, but the high pressures persisted.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hga1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|