(b)(4).This report is being submitted due to a retrospective review conducted under capa-(b)(4).On (b)(4) 2016, fse arrived at the site to address the reported event.Inspection of the device revealed that the fittings were restricting flow through the device.Fse corrected the fitting issue.The customer also reported that they were receiving low total areas on some samples.Further inspection revealed that the back cable was obstructing the sampler, resulting in a clot when the probe would go through the wash cycle.Fse replaced the cable to correct this issue.He subsequently ran twenty patient samples without errors.Final device pressure was stable at 8.89 mpa.No further action was required by field service.The most probable cause of the reported event was due to fault/ failure of internal tubing and fittings too tight.
|
On (b)(6) 2016, the customer reported ongoing issues with high pressure with their g7 analyzer.On (b)(4) 2016, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|