DEPUY SYNTHES PRODUCTS LLC SMALL BATTERY DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 532.010
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.A visual and functional assessment was performed which found that the device had low power, ran very slow and was sluggish.It was further observed that the device was not functioning with the epd-console.During service/repair, it was determined that there was damage insulation on the electronic control unit (ecu)-wire, the magnet from the control lever came out and the bearings and seals were worn out.The assignable root cause was determined to be due to normal wear out from use.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that during pre-surgery, it was discovered that the small battery drive device was not holding the charge well, had low power and was running sluggishly.During in-house engineering evaluation, it was observed that the magnet from the control lever came out.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMALL BATTERY DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7369231
MDR Text Key103417063
Report Number8030965-2018-52387
Device Sequence Number1
Product Code HWE
UDI-Device Identifier7611819110779
UDI-Public(01)7611819110779(11)011112
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532.010
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Manufacturer Received 03/02/2018
Initial Date FDA Received03/26/2018
Date Device Manufactured11/12/2001
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-