• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Physical Property Issue (3008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).This report is being submitted due to a retrospective review conducted under capa(b)(4).On (b)(6) 2017, an fse visited the customer to address the reported event.During servicing, fse found the instrument leaking around the t-handle and the nut on the line filter.They also found that the ferrule was not staying on the peek tubing.Fse then replaced the tube for the line filter.The instrument was verified as operational.There was no further action required by fse.The most probable cause of the reported event was the tube for the line filter.
 
Event Description
On (b)(6) 2017, a customer reported that a leak is occurring at the top of the filter housing where the outlet peek tubing attaches on their g8 analyzer.The customer stated that the leak occurred after they performed a prefilter replacement.Technical support tried troubleshooting with the customer in tightening the peek fitting but the leaking continued.A field service engineer (fse) was dispatched to address the reported event, which resulted in delayed reporting hga1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7372076
MDR Text Key103564135
Report Number8031673-2018-04400
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/26/2018
Distributor Facility Aware Date03/02/2017
Device Age2 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/26/2018
Initial Date Manufacturer Received 03/02/2017
Initial Date FDA Received03/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Congenital Anomaly; Death;
-
-