• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND RESONATE; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND RESONATE; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number G447
Device Problems Device Alarm System (1012); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/20/2018
Event Type  malfunction  
Manufacturer Narrative
No further investigation required.Information from field sufficient to make accurate reporting decisions.Should new information become available, this event will be further reviewed.
 
Event Description
Boston scientific received information that the patient with this device sought medical assistance due to beep tones.A latitude data request was filed to investigate for any system performance issues.The review illustrated high out of range left ventricular (lv) and right atrial (ra), lead pacing impedance measurements.The technical services consultant was able to confirm the tones were associated with an lv impedance greater than 2000 ohms.Patient data was suggestive of programming changes within the past few months due to the marked change in not only impedance but thresholds, at this timepoint.Ts recommended a programming change to set impedance alert notifications, at greater than 3000 ohms.The data review confirmed ra lead impedances measurement had been high since implant with the field representative confirming the system has no ra lead connected.No resulting adverse patient effects were reported and to date the device remains implanted and in service; no lv lead information was able to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7375523
MDR Text Key103676059
Report Number2124215-2018-05511
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/09/2019
Device Model NumberG447
Other Device ID NumberRESONATE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received02/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-