Boston scientific received information that the patient with this device sought medical assistance due to beep tones.A latitude data request was filed to investigate for any system performance issues.The review illustrated high out of range left ventricular (lv) and right atrial (ra), lead pacing impedance measurements.The technical services consultant was able to confirm the tones were associated with an lv impedance greater than 2000 ohms.Patient data was suggestive of programming changes within the past few months due to the marked change in not only impedance but thresholds, at this timepoint.Ts recommended a programming change to set impedance alert notifications, at greater than 3000 ohms.The data review confirmed ra lead impedances measurement had been high since implant with the field representative confirming the system has no ra lead connected.No resulting adverse patient effects were reported and to date the device remains implanted and in service; no lv lead information was able to be obtained.
|