Catalog Number 05.001.201
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2018
Event Type  malfunction  
Manufacturer Narrative
Reporter's phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during post-op, it was observed that the battery handpiece device was warming up.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device passed all operational specifications and no failure was identified.Therefore, the reported condition was not confirmed and an assignable root cause was not determined.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
The date device returned to manufacturer was documented as apr 04, 2018 in the initial report.This has been corrected to apr 03, 2018.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ   4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester 19380
6107195000
MDR Report Key7377458
MDR Text Key103692057
Report Number8030965-2018-52506
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819977815
UDI-Public(01)7611819977815(11)170222
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-