ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2018
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.External equipment was exchanged, however, the issue did not resolve.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed near the fantail, and damaged silicone overmold was observed on both top and bottom covers.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical test from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.The internal visual inspection noted silver migration across some electrical components.This device had moisture that exceeded the residual gas analysis limit.Leak testing did not reveal any leaks.Corrective actions were implemented.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7379599
MDR Text Key103925068
Report Number3006556115-2018-00137
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 03/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/30/2006
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/28/2018
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Supplement Dates FDA Received12/21/2018
01/16/2019
02/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2004
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age41 YR
-
-