Model Number G7
Device Problem Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/11/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This report is being submitted due to a retrospective review conducted under capa (b)(4).Device by manufacturer: on (b)(4) 2015, a technical support specialist (tss) addressed the reported event where the customer informed that they had failed api survey.Tss sent a new column and calibrators for evaluation.Customer advised new column and calibrators resolved the unacceptable pt results.No further action required.The probable cause of the event was lot to lot variability.
 
Event Description
On (b)(6) 2015, a customer reported failed api survey with their g7 analyzer.Customer reports failed api survey on the g7.Gly-11 13.3 [9.7-14.7], gly 12 - 7.9 [4.6-7.0], gly 13 - 13.1 [7.3-11.1], gly 14 - 8.3 [4.0-7.0], gly 15 - 16.8 [9.5-14.4].Calibrated with lot zs4002, 1.1129x + 0.6655.(b)(4), fc 36.Reports that quality control (qc) mean for mas dbcl 1702 in sept was 6.2 and 9.8.L1 qc mean for oct was 6.4, l2 was still 9.8.Tss asked to see qc api report.Advised to perform the monthly monitoring check and recalibrate.Monthly maintenance checks - check dilution process - 6.3 /wash; 6.1/di.Check sampling process - 6.1 whole blood; 6.3 manual dilution.The customer is unable to run hba1c patient samples.A technical support specialist (tss) addressed the reported event, which resulted in delayed reporting hba1c patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7
Type of Device
G7
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6506368123
MDR Report Key7380095
MDR Text Key103972230
Report Number8031673-2018-04139
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071132
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG7
Device Catalogue NumberG7: 019327
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/28/2018
Distributor Facility Aware Date11/11/2015
Device Age7 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer03/28/2018
Date Manufacturer Received11/11/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-