(b)(4) is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number (b)(4).Device evaluation by manufacturer: on (b)(6) 2018, fse was at customer's site to address the reported issue.Fse verified bent needle and replaced needle, and performed needle alignment.Fse cleaned needle wash block and orings.Fse completed sample precision test and ran quality control (qc) successfully and within acceptable range.Customer confirms the g8 instrument functioned as intended after completing the repair.A 13 month complaint history review and service history review for similar complaints was performed for the (b)(4) from (b)(6) 2017 through aware date (b)(6) 2018.There were two (2) similar complaints identified during the searched period, which includes this event.Sample needle the g8 variant operator's manual under section 5 - sampling needle replacement provides detailed instruction on how to replace the sample needle.The most probable cause of the reported event was due to bad sample needle alignment.
|
On (b)(6) 2018, customer reported they had a bent sample needle on the g8 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|