On (b)(6) 2015, a customer reported that the tubing on the left side of the column kept popping out , the br 33880 is recovering low, and the pressure reading was displayed higher than expected while in standby mode on the g7 instrument.Technical support assisted the customer with reinstalling the column and advised the customer to adjust the flow factor.The customer is unable to run patient samples on hba1c diabetes assay.On (b)(4) 2015, a field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hba1c.There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|